Instrument Configuration

You can personalize your Super-Fast Thermapen by changing several factory default set-

tings including the temperature display units from °F to °C, changing the temperature display

resolution from 1° to 0.1°, and disabling the auto-off feature. To reconfigure these settings,

open the battery compartment (see Replacing the Batteries section for instructions on open-

ing the battery compartment) and, using a bent paper clip, change the switch bank settings

as instructed below.

Switch 1 – Units

Changes the instrument’s display units between (°F) Fahrenheit

and (°C) Celsius. The factory default is °F or switch set furthest

from batteries. To change to °C, move the switch to the position

closest to the batteries.

Switch 2 – Resolution

Changes the instrument’s display resolution between 1°(F or C)

and 0.1°(F or C). The factory default is 1° resolution or switch set

farthest from the batteries. To change to 0.1° resolution, move the

switch to the position closest to the batteries.

Switch 3 – Auto-Off

Enables or disables the auto-off function. If the auto-off function is

enabled (factory default), the unit will automatically turn off after

10 minutes to preserve battery life. If disabled, the unit will remain

on until the probe is folded away and turned off by the user. The

factory default is auto-off enabled or switch set closest to the

batteries. To disable the auto-off function move the switch to the

position furthest from the batteries.

Switch 4 – Trim Adjust

If the Thermapen is within specification, we do not recommend

using the Trim Adjust. It should not be needed in normal use. This

mode allows the user to set an offset that will add or subtract up

to 3.6°F (2.0°C) to all readings of the Thermapen. It can therefore

be used to tune the calibration. For detailed instructions visit www.

thermoworks.com.

Calibration

The Thermapen is individually factory-calibrated using very high-precision temperature stan-

dards that are traceable to NIST (national standards). A certificate with actual test data is

supplied with your Thermapen. However, if you have any doubt about its accuracy, you can

check it yourself in a properly prepared ice bath. If it reads correctly in the ice bath, it is almost

certainly accurate at other temperatures.

Making a Proper Ice Bath

Fill a cup or glass with ice. Add just enough water to fill the gaps between the ice. (If the ice

is floating, the water beneath the ice will be at a higher temperature.) After a few moments,

immerse the tip of the Thermapen probe to the middle of the ice and gently stir. Within 3 or 4

seconds the Thermapen will read 32°F within 1°F. Do not let the tip of the probe rest against

an ice cube or it will read a lower temperature (it’s that fast and sensitive). If the reading is

within the published specification, the Thermapen is accurate. You can also test the Ther-

mapen in boiling water but you must first know the correct temperature of boiling water at

your altitude and at the current atmospheric pressure. A calculator can be found at