Sonics VCX750 (Serial No."Y through "AB")" User Manual
Page 19
RECALL: The recall function can retrieve from memory any of 9 stored control
parameters for verification or usage. To retrieve any parameters press the
SAVE/RECALL key.
The screen will display:
Press the ENTER/REVIEW key.
The screen will display:
Using the 1-9 numeric keys, enter the ID number.
The screen will display:
Press the ENTER/REVIEW key.
The screen will display:
Note: To review all the information that has been stored, press all the numeric keys
consecutively.
16
*RECALL JOB
SAVE JOB
R# T__: __: __
T__ __ °C
P _ _ _ _
E _ , _ _ _ , _ _ _ A _ _ _%
R4 T__: __: __
T__ __ °C
P _ _ _ _
E _ , _ _ _ , _ _ _ A _ _ _%
R4 TIME 5:30:25
TEMP __ __°C
PULSE 2.5 : 1.0
AMPL
40%
IMPORTANT
Proper care of the probe is essential for dependable operation. The intense
cavitation will, after a long period of time, cause the tip to erode, and the power
output to decrease without showing up on the wattmeter. The smoother and
shinier the tip, the more power will be transmitted into the sample. Any erosion
of the probe tip will increase the rate of future erosion. For that reason it is
recommended that after every 5 or 6 hours of use the tip be examined, and if
necessary, polished with emery cloth or an abrasive wheel, or machined in a
lathe. Since the probe is tuned to vibrate at a specific frequency, it is most
important that only the contaminated surface be removed. This procedure can
be repeated as long as the power monitor reading is less than 20% with the
probe out of the sample, when the AMPLITUDE control is set at 100. If the
reading exceeds 20%, the probe or replaceable tip should be replaced with a
new one.