TEMPERATURE: The temperature function prevents overheating of the sample by
continuously monitoring the sample temperature, and terminating the ultrasonics
when the temperature reaches a predetermined setpoint. The ultrasonics is
automatically reinstated when the temperature drops below the setpoint. If the
temperature of the sample must be monitored and/or controlled, insert the optional
temperature probe forcefully info the small jack on the rear panel, immerse the
temperature probe in the sample and press the TEMP key.

The screen will display:

Using the numeric keys set the high temperature limit (setpoint).

The screen will display:

Press the ENTER/REVIEW key.

The screen will display:

13

Probe Temperature 27°C

Temperature Setpoint __°C

Probe Temperature 27°C

Temperature Setpoint 35°C

TIME 5:30:25

TEMP

35°C

PULSE 2.5 : 1.0

AMPL

40%