Sonics VCX600 User Manual
Page 18
18
PULSE:
By inhibiting heat build- up in the sample, the pulse function enables safe
treatment of temperature sensitive samples at high intensity. In addition, pulsing
enhances processing by allowing the material to settle back under the probe after each
burst. The ON and OFF pulse duration can be set independently from .1 second to 9.9
seconds. During the OFF portion of the cycle, the red indicator on the
PULSE
key will
illuminate. If the OFF portion of the cycle exceeds two seconds, a cautionary message
-
CAUTION – PROBE ON STANDBY –
will warn the operator against touching the
ultrasonic probe. To set the pulser, press
PULSE
key.
The screen will display:
Using the numeric keys, set the ON portion of the cycle, and press the
ENTER/REVIEW
key.
The screen will display:
Using the numeric keys set the OFF portion of the cycle.
The screen will display:
Press the
ENTER/REVIEW
key.
The screen will display:
Pulse on __.__sec
Pulse off __.__sec
Pulse on 2.5 sec
Pulse off __.__sec
e.g.
Pulse on 2.5 sec
Pulse off 1.0 sec
e.g.
TIME 5:30:25 TEMP 35ºC
PULSE 2.5 : 1.0 AMPL 40 %