PULSE:

By inhibiting heat build- up in the sample, the pulse function enables safe

treatment of temperature sensitive samples at high intensity. In addition, pulsing
enhances processing by allowing the material to settle back under the probe after each
burst. The ON and OFF pulse duration can be set independently from .1 second to 9.9
seconds. During the OFF portion of the cycle, the red indicator on the

PULSE

key will

illuminate. If the OFF portion of the cycle exceeds two seconds, a cautionary message

-

CAUTION – PROBE ON STANDBY –

will warn the operator against touching the

ultrasonic probe. To set the pulser, press

PULSE

key.



The screen will display:



Using the numeric keys, set the ON portion of the cycle, and press the

ENTER/REVIEW

key.

The screen will display:



Using the numeric keys set the OFF portion of the cycle.

The screen will display:


Press the

ENTER/REVIEW

key.


The screen will display:


Pulse on __.__sec

Pulse off __.__sec

Pulse on 2.5 sec

Pulse off __.__sec

e.g.

Pulse on 2.5 sec

Pulse off 1.0 sec

e.g.

TIME 5:30:25 TEMP 35ºC
PULSE 2.5 : 1.0 AMPL 40 %