Sonics VCX1500 User Manual
Page 14
11
TEMPERATURE: The temperature function prevents overheating of the sample by
continuously monitoring the sample temperature, and terminating the ultrasonics
when the temperature reaches a predetermined setpoint. The ultrasonics is
automatically reinstated when the temperature drops below the setpoint. If the
temperature of the sample must be monitored and/or controlled, insert the optional
temperature probe forcefully info the small jack on the rear panel, immerse the
temperature probe in the sample and press the TEMP key.
The screen will display:
Using the numeric keys set the high temperature limit (setpoint).
The screen will display:
Press the ENTER/REVIEW key.
The screen will display:
Probe Temperature 27°C
Temperature Setpoint __°C
Probe Temperature 27°C
Temperature Setpoint 35°C
TIME 5:30:25
TEMP
35°C
PULSE 01 01
AMPL
40%