beautypg.com

Sonics VCX134AT User Manual

Page 10

background image

10

SECTION IV – USING THE ULTRASONIC ATOMIZER


AMPLITUDE:

Desired amplitude must be set in order for the Ultrasonic

Processor to be operational. AMPL. displays the amplitude selected e.g. 40%. To
set the amplitude at 40% when the ultrasonics is off, press the

AMPL

key and the

▲ ▼

key for a 40% reading on the screen, and then press the

ENTER/REVIEW

key.

The screen will display:

The Ultrasonic Processor is now ready for continuous operation. To energize the
ultrasonics, press the

START

key or the

footswitch. To de-energize the ultrasonics, press

the

STOP

key or release the footswitch.

To increase or decrease the amplitude in small increments when the ultrasonics is on,
depress the

ENTER/REVIEW

key twice to display AMPLITUDE CONTROL on the

screen, then depress the

or

key, as required.

CAUTION

Do not operate the ultrasonic power supply unless it is
connected to the converter

High voltage is present in the power supply – do not operate
with the cover removed.

AMPL 40 %

NOTE

To clear an erroneous entry press the

CLEAR

key.

NOTE

The

START

key and footswitch are mutually exclusive. If the process is initiated

by the

START

key, the footswitch becomes inoperative. If the process is initiated

by the footswitch, the

STOP

key becomes inoperative.

This manual is related to the following products: