PULSER:
By inhibiting heat build-up in the sample, the pulse function enables
safe treatment of temperature sensitive samples at high intensity. In addition,
pulsing enhances processing by allowing the material to settle back under the
probe after each burst. The ON and OFF pulse duration can be set independently
from 01 second to 59 seconds. During the OFF portion of the cycle, the red
indicator on the
PULSE
key will illuminate. If the OFF portion of the cycle
exceeds three seconds, a cautionary message
- Sonics in OFF Cycle -
will warn
the operator against touching the ultrasonic probe. To set the pulser, press
PULSE
key.
The screen will display:
Pulse on
--
sec
Pulse off
--
sec
Using the numeric keys, set the ON portion of the cycle, and press the ENTER/REVIEW
key.
The screen will display:
e.g.
Pulse on 01 sec
Pulse off
--
sec
Using the numeric keys set the OFF portion of the cycle.
The screen will display:
e.g.
Pulse on 01 sec
Pulse off 01 sec
Press the ENTER/REVIEW key.
The screen will display:
Off Cycle
On Cycle
Time 0:30:00
Pulse 01 01 Ample 40
%