8

5.9

Using The Timer

Using the incubator’s timer, you can run the shaker platform at a preset RPM for a preset
time. The timer can be set at intervals of one (1) minute up to a maximum of a 999 (16 hours
39 minutes).

Upon completion of the timing cycle, the TIMER alarm LIGHT will turn ON and an alarm will
sound.

You can interrupt the timer if you need to access the shaking platform before the timer
completes.

5.10

TO START A TIMED SHAKING PROCESS

Turn the TIMER switch to the ON (I) position.

Press the set timer up or down arrow key once. The display starts to blink off and on and
shows the current set time. Press the up or down arrow to select the desired time.

Approximately five seconds after you release the Up or Down arrow keys, the display stops
flashing and the timing interval begins.

5.11

To Interrupt A Timed Shaking Process

Turn the TIMER switch and the RPM switch to the OFF (O) position.

CAUTION

Wait for the mechanism to come to a complete stop before entering the chamber.

5.12

To Restart An Interrupted Timed Shaking Process

Turn the RPM and TIMER switch to the ON (I) position.