beautypg.com

Troubleshooting – Shellab SSI3-2 User Manual

Page 10

background image

10

TROUBLESHOOTING

TEMPERATURE

Temperature too high-display and
reference thermometer don’t match

1/ Controller set too high-see section 5.5.
2/ Controller failed on

– call Customer Service.

3/ Wiring error

– call Customer Service.

Display reads "HI"

Probe is defective, unplugged, or the wire is broken

– trace wire from

display to probe; move wire and watch display to see intermittent
problems.

Chamber temperature spikes over set
point and then settles to set point

Recalibrate

– see section 5.6.

Temperature too low-display and
reference thermometer don’t match

1/ OTP set too low

– turn OTP fully clockwise.

2/ Controller set too low

– see Section 5.5.

3/ Unit not recovered from door opening

– wait for display to stop

changing.
4/ Unit not recovered from power failure or being turned off

– incubators

will need 24 hours to warm up and stabilize.
5/ Element failure

– see if heating light is on; compare current draw to

data plate.
6/ controller failure

– confirm with front panel lights that controller is

calling for heat
7/ OTP failure

– confirm with front panel lights that OTP is operating

correctly.
8/ Loose connection

– check shadow box for loose connections.

Display reads "LO"

Sensor is plugged in backwards

– reverse sensor wires to controller.

Unit will not heat over a temperature that
is below set point

1/ Confirm that amperage and voltage match data plate
2/ Confirm that set point is set high enough

–turn OTP all the way

clockwise and see if heating light or safety light comes on.
3/ Check calibration

– using independent thermometer, follow

instructions in section 5.6.

Unit will not heat up at all

1/ Verify that controller is asking for heat by looking for controller light

if pilot light is not on continuously during initial start up, there is a
problem with the controller.
2/ Check amperage

– amperage should be virtually at maximum rated

(data plate) amperage.
3/ Do all controller functions work?
4/ Is the OTP set high enough?

– for diagnostics, should be fully

clockwise with the pilot light never on.
5/ Has the fuse/circuit breaker blown?

Indicated chamber temperature unstable

1/ ±0.1 may be normal.
2/ Is ambient radically changing

– either door opening or room airflow

from heaters or air conditioning ?

– stabilize ambient conditions.

Section

7

This manual is related to the following products: