Replacing the hepa filter assembly, Maintenance (continued), Hepa – Shellab SMO5CR-2 User Manual
Page 26: Eplacing the, Ilter, Ssembly

26 |
P a g e
MAINTENANCE (CONTINUED)
R
EPLACING THE
HEPA
F
ILTER
A
SSEMBLY
To replace the HEPA filter assembly, it is necessary to
remove or at least partly remove the door gasket. Replacing
the HEPA filter assembly is an opportune time to replace
the door gasket.
Note: Door removed in illustrations for clarity. Do not remove the
door from the oven.
1. Remove all shelves and shelf sliders from the oven.
2. Unscrew the nuts located at the top of the back wall of the
oven chamber. Remove the nuts and washers.
3. Remove the chamber gasket so that there is room to pull
the ceiling liner out through the door space. The gasket
may be fully or partially removed.
4. Remove the chamber ceiling liner by pulling it out through
the oven door space. The liner may require some side-to-
side motion to free up.
5. Remove the right wall air duct that the HEPA filter assembly
is mounted on by pulling it out through the door space.
Procedure continued on next page
1, 2, 3)
Remove
Shelves,
Nuts, and
Door Gasket
4) Ceiling
Liner
Removed
5) Right
Air Duct
and HEPA
Filter
Assembly
Removed
1
2
3
4