4

5

OPERATION

ASSEMBLY

Symbology

Changing the Batteries
Replace batteries when the low battery indicator

is displayed.

1. Turn rotary dial to OFF and disconnect the test

leads.

2. Unscrew and remove battery door.
3. Insert two (2) AAA batteries, according to the

polarity marked in
the battery com-
partment

4. Close the battery

door and tighten
screw securely.

WARNING

To avoid an electrical hazard, turn the rotary
dial to OFF and disconnect the test leads
before replacing batteries.

Read Operator’s Manual

Double Insulation

Risk of Electric Shock

Indicates that this instrument can
clamp on bare conductors when mea-
suring a voltage corresponding to the
applicable measurement category,
which is marked next to this symbol.

Earth (Ground)

Danger, Warning, or Caution

Battery

European Conformity Mark

Underwriters Laboratories, Inc.,
United States and Canada

1

Canadian Conformity Mark

Cat III

Classifi cation of transient overvolt-
ages, based on nominal line voltage
to earth.

DANGER

To avoid electrical shock:

Never make measurement on a circuit in
which voltage over 600 volts AC exists. Clamp
tips are designed not to short the circuit under
test. If equipment under test has exposed
conductive parts, however, extra precaution
should be taken to minimize the possibility
of shorting.
Do not use with the battery cover removed.
Disconnect the test leads from the instrument
for current measurement.

Before Use

Confi rm the rotary dial is set to the correct position,

the instrument is set to the correct measurement

mode, and the hold function is disabled. Otherwise,

desired measurement cannot be made.

LCD Backlight

The LCD backlight will turn off after about 10 min-

utes of inactivity. Push the backlight button to turn

the backlight on or off.

Making a Measurement

AC Current

WARNING

Only use MILWAUKEE test leads with the
MILWAUKEE
Clamp Meter.
Inspect test leads before each use. Use Clamp
Meter to run a continuity test.

1. Set the rotary dial to

position. AC mark is dis-

played.

2. Press the jaw opening trig-

ger to open the jaws and

clamp them around the

conductor under test. The

reading is displayed.

NOTE: Do not clamp over 2

or more wires at the same

time. Irregular results will

occur.

CAUTION

Maximum conductor size

is 1" diameter. During measurement,

keep the jaws fully closed to ensure accurate
measurements.

200A

DANGER

To avoid electrical shock:

Never make measurement on a circuit in
which voltage over 600 volts AC exists.
Do not use with the battery cover removed.
Keep fi ngers away from jaws during mea-
surements.

AC Voltage

1. Set the rotary dial to

position.

2. Connect the red test lead

to the VΩ terminal and the

black test lead to the COM

terminal.

3. Connect the test leads to

the circuit under test. The

reading is displayed.

1.5V

1.5V

1. Set the rotary dial to position.

2. Connect the red test lead to the VΩ terminal and

the black test lead to the COM terminal.

3. Connect the red test lead to the positive (+) side

and black test lead to the negative (-) side of the

circuit under test. The reading is displayed. A

reversed connection is indicated as a negative

value.

DANGER

To avoid electrical shock:

Never make measurement on a circuit in
which voltage over 600 volts DC exists.
Do not use with the battery cover removed.
Keep fi ngers away from jaws during mea-
surements.

DC Voltage

DANGER

To reduce the risk of electric shock for
Resistance and Continuity measurements,
never use the meter on an energized circuit.
If testing a capacitor, make sure it is fully
discharged before touching or attempting to
make a measurement.
Do not use with the battery cover removed.

Resistance

1. Set the rotary dial to posi-

tion.

2. Connect the red test lead

to the VΩ terminal and the

black test lead to the COM

terminal.

Confi rm “OL” is indicated on

the display, and then touch

the tips of the test leads to-

gether to short circuit them

to confi rm the zero indication.

3. Connect the test leads to

both ends of the resistor

under test. The reading is displayed.

CAUTION

After shorting the test leads, the displayed
value may not be zero due to the resistance
of test leads themselves.

Continuity

1. Set the rotary dial to

posi-

tion.

2. Connect the red test lead

to the VΩ terminal and the

black test lead to the COM

terminal.

Confi rm “OL” is indicated on

the display, and then touch

the tips of the test leads to-

gether to short circuit them

to confi rm the zero indica-

tion. A buzzer will sound.

3. Connect the test leads to

both ends of the conductor under test. If the

resistance under test is 30 Ω or less, the buzzer
will sound.

12V

Resistance/Continuity Measurements