LABEC M-MIA-LA User Manual
Page 95

Ultrasonic milkanalyser
Operation manual
04.11.09
93/107
pH Calibration 
pH Measuring 
pH En/Disable 
pH U Display 
------------------------- 
pH Test 
Co Meter Calibr 
Co Meter Test 
Co Meter En/Dis 
------------------------- 
Exit 
8. pH Calibration. 
Serves for calibrating the pH meter. For that purpose use 2 standard buffers, 
shown on the display as Low buffer (for example 5.00 pH) and High buffer 
(for example 7.00 pH). The procedure is following: 
 
1. Start menu Calibration. 
2. Put the pH probe in Low buffer. 
3. By pressing the buttons of the analyser enter the exact buffer value. The 
display shows: 
 
pH Calibr
Put Izopot buff
Buf=xx.xxx
4. Input the buffer value when the probe is in its isopotential point and with 
which our probes are accompanied with. 
Then the display shows: 
 
pH Calibr
Put Izopot buff
Buf=xx.xxx
V=x.xxxV Set
Where x.xxx is measured by the probe voltage. 
 
- AA13 (7 pages)
 - GW1060 (44 pages)
 - SS1 (4 pages)
 - AA-8BS (30 pages)
 - AA-22BS (32 pages)
 - AA-22N (32 pages)
 - AA-8S (33 pages)
 - AA-22S (26 pages)
 - Shinva-45B (21 pages)
 - AA9 (6 pages)
 - USC-5L (5 pages)
 - WS1.0 (3 pages)
 - BEM4 (4 pages)
 - LHF-40 (5 pages)
 - NBCT2 (4 pages)
 - NBCT7S (4 pages)
 - ICB-12 (3 pages)
 - LBC-4 (4 pages)
 - SB-SDC-6 (5 pages)
 - J-NBT (16 pages)
 - BOB1 (5 pages)
 - M-SOD-3636 (3 pages)
 - J-SWB35 (21 pages)
 - FXP16M (13 pages)
 - M-SI-100 (3 pages)
 - M-TMO1600 (3 pages)
 - M-HOG-020 (13 pages)
 - M-HOG010 (12 pages)
 - J-USRC (16 pages)
 - Z-L-450 (17 pages)
 - Z-L-450A (6 pages)
 - Z-L-530 (20 pages)
 - M-SCEN-206 (13 pages)
 - M-SCEN-207 (12 pages)
 - Z-ZHWY-2102 (9 pages)
 - GLF-90 (2 pages)
 - PLCF300 (2 pages)
 - H-DW-HW50 (12 pages)
 - DW-FL90 (8 pages)
 - B-IM0070FSC (5 pages)
 - GLR-70 (3 pages)
 - PLRGD-1400PT (3 pages)
 - PVR-130 (3 pages)
 - SPR-100 (3 pages)
 - CEMLS (2 pages)
 