LABEC M-MIA-LA User Manual
Page 39

Ultrasonic milkanalyser
Operation manual
04.11.09
37/107
Recalibrate 1
Put sample High
Cal meas=1/5
Follow the procedure till the 5th measurement.
After 5th measurement completion automatically appears the menu:
Cal 1 Samp Low
Fat = f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit
Which reminds to enter the values for the sample with low FAT? Enter the
values analogically to the procedure, described in the previous sample. After
entering the last parameter position the cursor on “Exit”, press “Enter” and
the following menu will appear:
Recal 1
Put sample Low 5
times
Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:
Recal 1
Put sample water
Which reminds for 5-times water measurement? There is no need to enter
values after placing the sample-holder. Measurement starts directly.
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)