5. Operation Procedure
(01) Insert power plug and switch power to “ON” at the right back side of centrifuge。
(02) Choose operation parameter and press【START】to begin operating. Meanwhile, the
light of【RUN】lighted up automatically.
(03) When there is a balance error, the light of【IMBAL】lighted up and stop operating.
Then press【OPEN】to check the rotor.
(04) Press【BRAKE】to set the time of brake.
(05) When operating time is up, the buzzer will sound automatically.
(06) During rotation, the【OPEN】button can not work.
(07) When the cover is open, the light of【DOOR】will light up and the【START】can
not work.
(08) Speed and time can be changed during operation, but it will not re-countdown.
(09) LED display time, speed and signal of breakdown.
(10) Emergency open hole can be used when electric power cut.
(11) The last setting parameter will be stored and display next time.
(12) Parameters will not be removed from memory storage when power cut off.
(13) To set speed, please refer to the form (Ⅱ-4).
Caution:
1. Please refer to operation panel, there are 10 sets of program available to store the desired
rpm and time. When you start to run the centrifuge next time, it will perform the last
running program. These programs don’t disappear at power off and on.
2. During rotation, if you want to change speed or time, then press【RPM】or【TIME】and
adjust ▲ or ▼ then press【RPM】or【TIME】to finish it separately.
3. During rotation, please don’t exceed the max. speed (refer page 2) of the centrifuge and
place substandard rotor or carrier. The centrifuge is without anti-explosion, so please
don’t use explosive material.