LABEC M-SCEN-206 User Manual
Page 11
![background image](/manuals/697880/11/background.png)
Repair methods:
Cause
Details of cause
Repairing way
Power
a. Power cord is not connected.
b. Power voltage is not proper.
Checking power plug, replace fuse
Lid problem Lid can not be closed
Contact local dealer
PC board Parts can not connect well
Contact local dealer
Motor
problem
Over- heat, carbon brush use out
Stop using it and wait for cool, replace carbon
brush or contact local dealer.
Imbalance Carrier or tube is put improperly. Please refer operation description
6. Signal of display:
Signal Status
Description
Repairing
Motor can not drive
Carbon brush connect badly.
Wire matched badly.
The outlet of controller is out of
order.
1. check carbon brush
2. check wiring
3. check the outlet of
controller
4. replace controller
Motor stops during
operation
Carbon brush connect badly.
Wire matched badly.
The outlet of controller is out of
order.
1. check carbon brush
2. check wiring
3. check the outlet of
controller
4. replace controller
Err-1
Sensor can not detect well
Sensor is out of order
Wire matched badly.
The outlet of controller is out of
order.
1. check sensor or replace new
one
2. check wiring
3. check the outlet of
controller
4. replace controller
Motor is over-heat
Temp. switch skips away.
After temp. is cooler then
operate it again.
Hot
Motor stops running
Temp. switch is out of order.
Wire matched badly.
1. check temp. switch or
replace new one.
2. check wiring.
9