beautypg.com

Setup menu – LABEC GW1060 User Manual

Page 22

background image

Rev. 00

GW2045 – GW1060 Manual

Pag. 21 di 43


SETUP MENU

SETUP menu, available functions:

1. Position 1: For dose of detergent in pump P1. Set the ml/litre required as dose of detergent with

maximum threshold allowed at 20ml/litre. In 1ml steps (increase/decrease by using buttons (5) and
(4)). P1 Default dosage is 5 ml/litre.

2. Position 2: Similar to Position 1 for P2, neutraliser. P2 Default dosage is 3 ml/litre.

3. Position 3: Similar to Position 1 for P3. P3 Default dosage is 0 ml/litre.

4. Position 4: Setting of hardness of water in °f, from 5 to 60 with 5°f steps. Default is a w ater

hardness of 40°f. Important parameter for the corre ct functioning of softener incorporated into
appliance.

5. Position 5: Choice for automatic opening of door at end of cycle. “ON” as default. ON and OFF

appear on the display in segments to select the two options.

6. Position 6: Printer language - to be selected from the various languages available; default is

English.

Selected Language

German

French

Spanish

Italian

English

Abbreviation on Display

dE

Fr

ES

It

En

7. Position 7: Demineralised water attachment, yes/no.

Demineralised water attachment Present Absent

Abbreviation on Display

dn

--

8. Position 8: Date-time (for details se paragraph DATE AND TIME SETUP).

It is possible to access and change the following parameters using programme selection and the
buttons (5) and (4):

Setup field

Year month day Time Minutes Seconds

Inactive

position

Inactive

position

Switch

position

1

2

3

4

5

6

7

8