10 service documentation, 1 overview, setting regulation, tolerances, Service documentation – KERN & SOHN KFD 1500V20LM User Manual
Page 14: Overview, setting regulation, tolerances
10 Service documentation
• This chapter is only intended for a balance specialist!
• At every corner of the weighing bridge a DMS weighing cell is installed.
• The analogue-digital transformation occurs in the display unit. Also all
the balance and country-specific data are stored there.
10.1 Overview, setting regulation, tolerances
Testing and setting regulations:
Capacity
600 kg
1500 kg
Readability
200 g
500 g
Min
4 kg
10 kg
Max
600 kg
1500 kg
1/3 corner load
200 kg
500 kg
Tolerance
200 g
500 g
KFD V20-IA-e-1110
14