3 output log (continuous output), 4 remote control instructions – KERN & SOHN KFN-TM User Manual
Page 49

KFB/KFN-TM-BA_IA-e-1323
49
Symbols:
ST
Stable value
US
Instable value
GS / GW
Gross weight
NT
Net weight
TW
Tare weight
NO
Number weighing processes
TOTAL
Total of all individual weighings
Space line
Space line
• Counting
10.3 Output log (continuous output)
• Weighing
HEADER1: ST=STABLE
, US=UNSTABLE
HEADER2: NT=NET
, GS=GROSS
10.4 Remote control instructions
Command Function
Printout examples
S
Stable weighing value for the weight is
sent via the RS232 interface
ST,GS 1.000KG
W
Weighing value for the weight (stable or
unstable) is sent via the RS232 interface
US,GS 1.342KG
ST,GS 1.000KG
T
No data are sent, the balance carries out
the tare function.
-
Z
No data are sent, the zero-display
appears.
-
P
Quantity will be sent via the RS232-
interface
10PCS
****************************
PCS
100
****************************