KERN & SOHN KFN-TM User Manual
Page 17

KFB/KFN-TM-BA_IA-e-1323
17
How to carry out an adjustment:
Confirm menu setting „noLin“ by
.
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press
.
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with
.
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press
.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.