7.4 Plus/minus weighings
Operation
Display
Switch on balance by pressing
and wait for the
[0] display.
Put the nominal weight on the weighing plate and tare
to [0] using the
button.
(Example)
Remove the nominal weight.
The nominal weight is displayed as negative display.
Put the test objects subsequently on the weighing
plate, the respective deviation from the nominal weight
is displayed with the respective sign to „+“ and „-“.
(Example)
Back to weighing mode by pressing the