Explanation of the data transfer – KERN & SOHN EMB 500-1BE User Manual
Page 45

EMB-V-BA-e-1422
45
3. Explanation of the data transfer
rE Cr:
PRINT key
Press the PRINT key, at stable weight is transferred.
Remote control commands
The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.
Take into account that the following remote control commands must be sent without
a subsequent CR LF.
s
Function:
Stable weighing value for the weight is sent via the RS232
interface
w
Function:
Weighing value for the weight (stable or unstable) is sent
via the RS232 interface
t
Function:
No data are sent, the balance carries out the tare function.
a. Format for stable values
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17
18
M
S
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B U
1
U
2
U
3
CR
LF
b. Format in case of fault
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17
18
B B B B B B B B B B B E
r
r
o
r
CR
LF
c. Format for unstable values
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17
18
M
S
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B B B B CR
LF
AU PC:
The weighing values are sent automatically and continuously, no matter if the value
is stable or unstable.
d. Format for stable values
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17
18
M
S
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B
U
1
U
2
U
3
CR
LF
e. Format in case of fault
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17
18
B B B B B B B B B B B E
r
r
o
r
CR
LF
f. Format for unstable values
1
2
3
4
5
6
7
8
9
10 11 12
13 14 15 16 17
18
M
S
N
1
N
2
N
3
N
4
N
5
N
6
N
7
N
8
N
9
N
10
B B B B CR
LF