beautypg.com

Casio 5214 User Manual

Page 7

background image

Operation Guide 5114 5214

E-48

Barometer/Thermometer Mode Second Hand Units

The second hand indicates barometric
pressure differentials up to ±10hPa (in
1-hPa units).

The nearby illustration, for example,

shows what the second hand would
indicate when the calculated differential
is approximately 5 hPa (approximately
0.15 inHg).

The second hand will point to OVER or

UNDER if the barometric pressure
differential is outside the allowable
range of the scale.

The second hand will move to 9 o’clock

if a sensor reading could not be taken
for some reason or if the reading is
outside the allowable range.

Barometric pressure is calculated and

displayed using hPa as the standard.
The barometric pressure differential
also can be read in inHg units as shown
in the illustration (1 hPa = 0.03 inHg).

−0.15

0.15

−5

5

0

0

−10

10

−0.3

0.3

Second hand

inHg values

hPa values

Current pressure

greater than most

recent measured

pressure

Current pressure

less than most

recent measured

pressure

−0.15

0.15

−5

5

0

0

−10

10

−0.3

0.3

Second hand

inHg values

hPa values

Current pressure

greater than most

recent measured

pressure

Current pressure

less than most

recent measured

pressure

E-49

Pressure Sensor and Temperature Sensor Calibration

The pressure sensor and temperature sensor built into the watch are calibrated at the factory and normally
require no further adjustment. If you notice serious errors in the pressure readings and temperature
readings produced by the watch, you can calibrate a sensor to the reading of another device to correct the
errors.

Important!

Incorrectly calibrating the pressure sensor can result in incorrect readings. Before changing the

calibration value, make sure the readings you are using are reliable and accurate.

Incorrectly calibrating the temperature sensor can result in incorrect readings.

Carefully read the following before doing anything.

Compare the readings produced by the watch with those of another reliable and accurate thermometer.

– If adjustment is required, remove the watch from your wrist and wait for 20 or 30 minutes to give the

temperature of the watch time to stabilize.

E-50

To calibrate the pressure sensor and the temperature sensor

1. Take a reading with another measurement device to determine the

exact current barometric pressure or temperature.

2. With the watch in the Timekeeping Mode or in any of the sensor

modes, press

B to enter the Barometer/Thermometer Mode.

3. Hold down

E until each of the following steps occurs.

Hold SET appears on the digital display.

➝ The current temperature

value fl ashes.

4. Press

D to move the fl ashing between the temperature value and

barometric pressure value, to select the one you want to calibrate.

5. Use

A (+) and C (–) to adjust the calibration value in the units shown

below.

Temperature 0.1 °C (0.2 °F)
Barometric Pressure 1 hPa (0.05 inHg)

To return the currently fl ashing value to its initial factory default

setting, press

A and C at the same time. OFF will appear at

the fl ashing location for about one second, followed by the initial
default value.

6. Press

E to return to the Barometer/Thermometer Mode screen.

Press

D.

Press

D.

E-51

Barometer and Thermometer Precautions

The pressure sensor built into this watch measures changes in air pressure, which you can then apply

to your own weather predictions. It is not intended for use as a precision instrument in offi cial weather
prediction or reporting applications.

Sudden temperature changes can affect pressure sensor readings.

Temperature measurements are affected by your body temperature (while you are wearing the watch),

direct sunlight, and moisture. To achieve a more accurate temperature measurement, remove the
watch from your wrist, place it in a well ventilated location out of direct sunlight, and wipe all moisture
from the case. It takes approximately 20 to 30 minutes for the case of the watch to reach the actual
surrounding temperature.

E-52

Taking Altitude Readings

The watch displays altitude values based on air pressure readings taken by a built-in pressure sensor.

How the Altimeter Measures Altitude

The altimeter can measure altitude based on its own preset values (initial default method) or using a
reference altitude specifi ed by you.

When you measure altitude based on preset values
Data produced by the watch’s barometric pressure sensor is converted to approximate altitude based on
ISA (International Standard Atmosphere) conversion values stored in watch memory.

When you measure altitude using a reference altitude specifi ed by you
After you specify a reference altitude, the watch uses that value to convert
barometric pressure readings to altitude (page E-62).

When mountain climbing, you can specify a reference altitude value in

accordance with a marker along the way or altitude information from a
map. After that, the altitude readings produced by the watch will be more
accurate than they would without a reference altitude value.

A

B

A

B

E-53

To take an altimeter reading

1. Make sure the watch is in the Timekeeping Mode or any one of the

sensor modes.

The sensor modes are: Compass Mode, Barometer/Thermometer

Mode, and Altimeter Mode.

2. Press

A to start Altimeter measurement.

When you press

A, the second hand will fi rst point in the direction of

the

A (ALTI) button. This indicates the Altimeter Mode. After that,

the second hand will point at the current altitude differential.

ALTI will appear on the digital display, indicating that Altimeter
measurement is in progress. The fi rst reading will appear on the
digital display after about four or fi ve seconds.

The current altitude value is displayed in units of 5 meters (20 feet).

After the fi rst reading is obtained, the watch continues to take

altimeter readings automatically every fi ve seconds for the fi rst
three minutes, and then every two minutes after that (under initial
default settings).

If you leave the watch in the Altimeter Mode, it will update the displayed altitude value regularly and

indicate reading-to-reading changes in graph form.

You can use the procedure under “Selecting an Altitude Auto Measurement Method” (page E-55) to

specify the altitude auto measurement method you want to use.

3. After you are fi nished using the Altimeter, press

D to return to the Timekeeping Mode and stop auto

measurement.

The watch will return to the Timekeeping Mode automatically if you do not perform any operation for

about 10 hours after entering the Altimeter Mode (under initial default settings).

Altitude differential
indicated by second hand

Altitude
graph

Current
attitude

Altitude differential
indicated by second hand

Altitude
graph

Current
attitude

E-54

Reading the Altitude Graph

The altitude graph shows Altimeter Mode auto measurement readings over time.

The vertical axis of the graph represents altitude, and each dot stands

for 10 meters (40 feet).

The horizontal axis represents time. For the altitude readings taken

during the fi rst three minutes after you start an altimeter measurement
operation, each dot represents fi ve seconds. After that, each dot
represents two minutes (under initial default settings).

An out of range reading or a measurement error will cause the column

of dots for that reading to be blank (skipped).

Note

The measurement range for altitude is –700 to 10,000 meters (–2,300 to 32,800 feet).

The displayed altitude value changes to - - - - if an altitude reading falls outside the measurement

range. An altitude value will reappear as soon as the altitude reading is within the allowable range.

Normally, displayed altitude values are based on the watch’s preset conversion values. You also can
specify a reference altitude value, if you want. See “Specifying a Reference Altitude Value” (page E-61).

You can change the unit for displayed altitude values to either meters (m) or feet (ft). See “To specify

temperature, barometric pressure, and altitude units” (page E-69).

Altitude

Latest
reading

Time

Altitude

Latest
reading

Time

E-55

Selecting an Altitude Auto Measurement Method

You can select either of the following two altitude auto measurement methods.
0’05 1H : Readings at fi ve-second intervals for one hour
2’00 10H : Readings at fi ve-second intervals for the fi rst three minutes followed by two-minute

intervals for approximately 10 hours

Note
If you do not perform any button operation while in the Altimeter Mode, the watch will return to the
Timekeeping Mode automatically after 10 hours (altitude auto measurement method: 2’00 10H ) or after
one hour (altitude auto measurement method: 0’05 1H ).

To select the altitude auto measurement method

1. In the Altimeter Mode, hold down

E until each of the following steps

occurs.

Hold SET appears on the digital display.

Hold SET disappears.

2. Press

D to display the current altitude auto measurement method

setting.

This will cause either 0’05 or 2’00 to fl ash on the digital display.

3. Press

A to toggle the altitude auto measurement method setting

between 0’05 1H and 2’00 10H .

4. Press

E to exit the setting screen.

This manual is related to the following products: