Grant Instruments TBT-T100IN User Manual
Page 18

31082 V3
TBT-T100IN
Page 17
Operating Manual
www.grantinstruments.com
Display shows 
“Open” 
Faulty temperature 
probe 
Have a competent person check the probe for an 
open circuit fault or contact Grant. 
Display shows 
“Shrt” 
Faulty temperature 
probe 
Have a competent person check the probe for a 
short circuit fault or contact Grant. 
Unit showing 
erratic 
temperatures 
Calibration values not 
set correctly 
Restore the factory calibration settings (see section 
5.2.7) then re-calibrate if required (see section 
5.2.6). 
Display shows 
“Hot” 
Heating water at or 
near to boiling without 
lid 
 
Heating very large 
volumes of liquid with 
large thermal losses 
Add a lid to reduce thermal losses and leave unit 
r
unning with the “Hot” warning present. If the alarm
has not cleared within 1 hour contact Grant. 
 
Add measures to reduce thermal losses and leave 
unit running with the “Hot” warning present. If the 
alarm has not cleared within 1 hour contact Grant. 
Display shows 
“Fuse” 
Faulty fuse(s)
Have a competent person check the internal fuses or 
contact Grant. 
Display shows 
“Crct” 
PCB fault
Have a competent person check the product or 
contact Grant. 
Display shows 
“Self” 
Safety relay fault
Have a competent person check the product or 
contact Grant. 
Stirrer motor not 
rotating 
Stirring propeller or 
pump impeller is 
obstructed 
 
Faulty motor 
Clear obstruction. 
 
 
 
Have a competent person check the motor or contact 
Grant. 
Display shows 
“Cold” 
Accidentally entered 
manufacturer reset 
mode 
Warning can be cleared by pressing S button, but 
note that all settings (including any factory 
calibration) will be reset to the default values. 
Contact Grant Instruments service department for 
assistance if required. 
 
For any other errors or service requests, please contact Grant Instruments service 
department. 
 