0 technical specifications, Operating conditions, Electrical details – Grant Instruments LTC2 User Manual
Page 18: Performance
30933 V2
LTC2
Page 17
Operating Manual
www.grantinstruments.com
6.0 Technical specifications
6.1
Operating conditions
Ambient temperature range
5 to 40°C
Altitude above sea level
Up to 2,000m (6,500ft)
Operating environment
Indoor use only
Maximum relative humidity
80% RH up to 31°C decreasing to 50% RH at 40°C
6.2
Electrical details
Mains supply:
220-240V @ 50Hz or 110-120V @ 60Hz
Pollution degree:
2
Installation category:
II
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3
Performance
Specification
LTC2
LTC2L
Typical cooling power at an
ambient of 20°C
@ 20°C
250W
@ 0°C
140W
@ -10°C
100W
@ -20°C
35W
Stability (DIN 12876)
±0.1°C †
Uniformity (DIN 12876)
±0.1°C
Settable temperature range
-20°C to 120°C
Working temperature range
-20°C to 100°C
Refrigerant R134a
140gm
Tank capacity
5L
Top opening
110/145mm
Max pump head pressure
210mBar
Max pump flow rate
16L/min
Safety 100°C limit
Cut-out
Safety H
2
O freezing protection
5°C thermostat & switch
Max current consumption (both units)
8.0A
15.5A*
Heater power
1.3kW
1.4kW
*
Requires two separate mains cords
† Temperature stability may be affected (±0.3C) in the presence of strong RF fields (10V/m) at
380-400MHz. This level of interference is very unlikely and is only likely to be encountered in
electrically noisy industrial locations as defined in EN61326)