5.0 Operating procedures
5.1
Operation
5.1.1 Filling the tank
Fill the tank with approximately 20 litres of water. The T100 has low level float switch
which will alarm if the liquid level drops below the minimum required level and the unit
will switch off the heater and stop temperature control.
5.1.3 Emptying the tank
The MST20 should be emptied to a safe level prior to moving.
Allow the liquid temperature to fall to a safe level before emptying. Take reasonable
precautions to prevent accidental spillage.
5.1.4 Setting up and switching on
Connect the T100 to a grounded (earthed) electrical power supply with voltage and
frequency within the range specified on the serial number plate.
Attach the T100 securely to the required bath or vessel. Add the appropriate working
liquid to the bath to at least the minimum recommended fill level such that the float level
switch is fully raised.
Switch on the T100 using the switch on the rear of the unit. The motor will start
immediately and the buzzer will sound while the unit starts up. During start up the
display will show the model type, followed by the software version and finally the current
liquid temperature. The T100 is ready to use.
To disconnect the equipment from the mains supply, remove the mains plug from the
mains supply outlet.