GLS Aqua Plus

29339 Ver. 3

Operating Manual

Page 12

www.grant.co.uk

EN

7. Press the enter key to confirm the value and the bath will display „

‟. Wait for the

bath to return to normal mode.


High calibration point (

)

1. Set the bath to a temperature between 70°C and 90°C, allow the bath to stabilise,

indicated by the display not fluctuating by more than 0.2°C.

2. Measure the temperature of the water with a calibrated thermometer and make a

note of this reading.

3. Press the enter key. The display will show „

‟.

4. Use the „+‟ or „-‟ keys to navigate to „

‟ and press the enter key.

5. Use the „+‟ or „-‟ keys to navigate to „

‟ and press the enter key.

6. Using the „+‟ or „-‟ keys adjust the display until the value display is the same as the

reading taken earlier.

7. Press the enter key to confirm the value and the bath will display „

‟. Wait for the

bath to return to normal mode.


Setting the calibration (

)

1. Press the enter key. The display will show „

‟.

2. Use the „+‟ or „-‟ keys to navigate to „

‟ and press the enter key.

3. Use the „+‟ or „-‟ keys to navigate to

‟ and press the enter key to apply the new

low and high calibration values.
The bath is now calibrated and the display will read the corrected value.


Restoring factory calibration (

)

1. Press the enter key. The display will show „

‟.

2. Use the „+‟ or „-‟ keys to navigate to „

‟ and press the enter key.

3. Use the „+‟ or „-‟ keys to navigate to

‟ and press the enter key to restore the

factory settings.