Grant Instruments CS200G User Manual
Page 4
OLS200, CSG200
15214 Ver 9 - January 2010 57U
Operating Manual
Page 3
www.grant.co.uk
2
Getting Started
2.1
Unpacking
Remove the packing materials carefully, and retain for future shipment or storage.
2.1.1
Shaking bath - OLS200
The pack should contain:
shaking bath
trolley
drive magnet
drain insert (located in a clip on the rear panel of OLS200)
mains cable
these instructions.
2.1.2
Universal Tray
Tray fitted with springs for use with flasks.
2.1.3
Test Tube Tray
Tray for use with test tube racks.
2.1.4
Cooler - CS200G
Refrigerated immersion cooler, for operating temperatures down to 0°C.
2.1.5
Cooling coil - CW200
For operating temperatures at or below ambient, a Grant CW200 heat exchange coil can be used
to maintain the temperature at 2°C above the temperature of the coolant.
2.2
Installation
2.2.1
Location
Position the bath on a firm level surface with the controls facing the operator and adjust the
position of the feet so that the bath stands level. This can be done after filling with liquid to the top
of the trolley, and levelling the water against the top of the trolley as a reference. It is necessary
that the bath is level in order to achieve optimum performance.
2.2.2
for operating temperatures below ambient
The unit is insulated for use at temperatures below ambient. It is recommended that the gabled lid
is used to reduce heat gain.
To fit the cooling coil: remove the flask tray and trolley, then remove the base tray by undoing the
retaining 'N' nuts (see figures 1 and 5).
CS200G: Stand the CS200G next to the OLS 200 and position the cooling coil in the bottom of the
tank, ensuring that it lies flat, with the feed pipe at the rear.
CW200: Place the CW200 cooling coil in the bottom of the tank, ensuring that it lies flat, with the
input and output pipes at the rear.
To secure the cooling coil in position, replace the base tray and washers and tighten the four
retaining 'N' nuts (see figure 5).
Ensure that the CS200G is located so that the ventilation grilles are not less than 100mm from
any obstruction.
To use the lid: remove the slot cover and replace in the alternative position, to allow the
connecting pipe to pass through.
2.2.3
for operating temperatures above 60°C
At operating temperatures above 60°C a lid MUST be fitted. The Grant sloping stainless steel lid
directs condensate into the bath to prevent it from dripping into open flasks or tubes.