Safety
The following symbols marked on the equipment mean:-
Caution: Read these operating instructions fully before use and pay particular attention to sections
containing this symbol
Caution: Surfaces can become hot during use.
Always observe the following safety precautions
Use only as specified by the operating instructions or the intrinsic protection may be impaired.
After transport or storage in humid conditions, dry out the unit before connecting it to the supply
voltage. During drying out the intrinsic protection may be impaired.
Connect only to a power supply with a voltage corresponding to that on the serial number label.
Connect only to power supply which provides a safety earth (ground) terminal.
Before moving, disconnect at the power supply socket. Do not remove the IEC connector.
Do not check the temperature by touch; use the temperature display or a thermometer.
To reduce the risk of eye injury during high temperature operation, use safety goggles or spectacles.
Do not touch surfaces which become hot during high temperature operation.
Ensure that the operating temperature is less than the maximum operating temperature of your
sample material.
Set the adjustable overtemperature cut-out after setting or changing the set temperature, and reset it
at monthly intervals to check that the cut-out is operating correctly.
Ensure that the mains switch is easily accessible during use.
Do not block or restrict ventilation slots.
If liquid is spilt inside the unit, disconnect it from the power supply and have it checked by a
competent person.
It is the user’s responsibility to carry out appropriate decontamination if hazardous material is spilt on
or inside the equipment.
Do not connect to a power supply or switch on before filling the tank.
Take care when topping up or draining, as the liquid in the tank may be very hot.
If the alarm lamp is illuminated do not touch the liquid or the heater, they may be very hot. Refill
carefully, a hot heater can cause a spattering of very hot water droplets and scalding steam.
Always use a lid when operating above 60°C. Take care when rising and removing the lid, it may be
hot. Steam and hot vapours can cause scalding.
Drain before moving the bath. Before draining allow the liquid to cool to below 50°C or heat to above
10°C.
Use only liquids specified in the operating instructions, within the specified temperature range. If the
alarm lamp is illuminated the liquid temperature may be above its recommended maximum: do not
inhale the vapours given off as they may be toxic; liquids should be safely discarded and replaced.