beautypg.com

Grant Instruments CS200G User Manual

Page 11

background image


OLS200, CSG200

15214 Ver 9 - January 2010 57U

Operating Manual

Page 10

www.grant.co.uk

6

Technical specification

This equipment is designed for indoor use in laboratory conditions, with room temperature
between 5°C and 40°C, and 80% relative humidity up to 31°C. Performance figures apply in
ambient temperature between 10°C and 35°C.

Installation category II (Transient voltages). Pollution degree 2 in accordance with IEC 664. For
operation at altitudes up to 2000 metres.

As Grant Instruments is committed to a continuous programme of improvement, specifications
may be changed without notice.



6.1

OLS200 orbital/linear shaking bath

Temperature range

0 to 99°C

Stability (DIN 58966)

± 0.1°C

Uniformity

± 0.1°C

Temperature setting/display

digital/LED

Display resolution

0.1

Shaking speed range

20 to 200 rpm

40 to 360 strokes/min

Linear - stroke length

18, 28, 36 mm

Orbital radius

9mm

Shaking speed setting/display

digital/LED

Display resolution

1

Heater power

230V

– 1450VA

115V

– 1200VA

Power rating

230V

– 1500VA

115V

– 1250VA

Tank dimensions l/w/d

505/300/200 mm

Shaking tray area

375/235 mm

Flask immersion min./max.

0/90 mm

Liquid depth, in the tank

40mm to 130mm

Overall dimensions l/w/d

555/325/300 mm

Mass

15kg

Heat up time from 25°C to 70°C

32 minutes

with water and the lid on

Safety: overtemperature/

adjustable cut-out

low liquid level

Supply voltage range

230V

10% @

50/60Hz


120V

10% @

50/60Hz

This manual is related to the following products: