Defibtech DDU-120 Series User Manual
Page 17

9
DAC-530E-EN-BH
WARNING
In patients with cardiac pacemakers, the DDU-120 AED may have reduced 
sensitivity and not detect all shockable rhythms. If you know the patient has an 
implanted pacemaker, do not place electrodes directly over an implanted device.
2.2.4 Shock Delivery
WARNING
Do not allow defibrillation pads to touch each other, or to touch other ECG 
electrodes, lead wires, dressings, transdermal patches, etc. Such contact 
can cause electrical arcing and patient skin burns during defibrillation and may 
divert defibrillating energy away from the heart.
WARNING
During defibrillation, air pockets between the skin and defibrillation pads can 
cause patient skin burns. To help prevent air pockets, make sure self-adhesive 
defibrillation pads completely adhere to the skin. Do not use dried out or 
expired defibrillation pads.
2.2.5 Maintenance
WARNING
Periodic user-initiated and automatic self-tests are designed to assess the 
DDU-120 AED’s readiness for use. However, no degree of testing can assure 
performance or detect abuse, damage, or a defect that occurred after the most 
recent test is completed.
WARNING
Use of damaged equipment or accessories may cause the device to perform 
improperly and/or result in injury to the patient or operator.
CAUTION
Improper maintenance can cause the DDU-120 AED not to function. Maintain 
the DDU-120 AED only as described in this User Manual. The AED contains 
no user serviceable parts – do not take the unit apart.
WARNING
No modification of this equipment is allowed.
