12 Electromagnetic Conformity
12.1 Guidance And Manufacturer’s Declaration
The essential performance of the DDU-2000 Series AED is successful delivery of defibrillation therapy and accurate
differentiation between shockable and nonshockable rhythms.
DDU-2000 Series AEDs are intended for use within the electromagnetic environment specified below. The customer
or the user of the DDU-2000 Series AED should assure that it is used within the stated environmental specifications.
ELECTROMAGNETIC EMISSIONS
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
The DDU-2000 Series AED uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
CISPR 11
Group 1
Class B
CISPR 22
Class B
FCC Part 15
Class B
Harmonic emissions IEC 61000-3-2
Not applicable
Battery operated equipment
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Battery operated equipment
ELECTROMAGNETIC IMMUNITY
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
There are no special requirements
with respect to electrostatic discharge.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for
power line
supply lines
±1 kV for
input/output
lines
Not applicable
Battery operated equipment
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
Not applicable
Battery operated equipment
Voltage dips, short interruptions and
voltage variations on power supply
input lines IEC 61000-4-11
Not applicable Not applicable
Battery operated equipment
Power frequency (50/60 Hz) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should not be greater than levels
characteristic of a typical location in a
commercial or hospital environment.