beautypg.com

7 repair, Maintenance and tr oubleshooting – Defibtech DDU-2450 Series User Manual User Manual

Page 39

background image

Symptom

Possible Cause

Corrective Action

“ Apply pads to patient’s bare chest as

shown” prompt

Pads not connected to patient

Place pads on patient

Pads not making good
connection to patient

Check pad connection to
patient

Pads or pad cable damaged

Replace pads

“Poor pad contact to patient” or
“Press pads firmly” prompt

Dry pads

Replace pads

Partial pad connection

Check that pads are
placed securely on
patient

“Check pads” prompt

Pads touching

Separate pads and place
correctly on patient

“Stop motion” prompt

Patient motion has
been detected

Stop patient motion

“ Stop interference” prompt

External interference
has been detected

Stop external
interference

“ Analyzing interrupted” prompt

Motion or interference detected

Stop motion or
interference

“Shock cancelled” prompt

Patient’s ECG rhythm changed

No action necessary

Shock button not pressed within
30 seconds

Press shock button
within 30 seconds

Low battery – insufficient to
charge

Replace battery pack

Bad pad to patient connection

Check that pads are
placed securely on
patient

Dry pads

Replace pads

“ Replace data card” prompt

DDC card is full

Replace DDC card with a
card that is not full

DDC card has failed

Replace DDC card

“ Pads missing” prompt

Pads not connected to unit

Make sure pads
connector is oriented
correctly and fully
inserted into unit

7.7 Repair

The DDU-2000 Series AED contains no user serviceable parts. If the unit needs servicing, call Defibtech.
(Refer to

Chapter 14

of this manual for contact information.)

Troubleshooting (continued)

7.

Maintenance

And

Tr

oubleshooting

39

DAC-E2510EN-BB

This manual is related to the following products: