8.0 Maintenance and Service
All Boekel laboratory products are designed to comply with IEC1010-1.
No routine maintenance is required.
8.1
Cleaning
Disengage power cord prior to cleaning. If a spill occurs, use appropriate
clean-up procedures as required for radiation or biohazard control. The
outer casing can be cleaned with a cloth dampened with water. Do not
immerse the incubator in water.
8.2
Replacement of Fuses (all models except 115 V 5.0 Cu. Ft.)
Caution – Double pole / Neutral Fusing
There are two supply fuses located in the fuse drawer. To change the
fuses:
• Turn power switch to the off (O) position
• Disconnect the unit from the power supply
• Remove the line cord from the power entry module on the back of the
unit
• Pull back on the fuse drawer catch
• Pull out the fuse drawer (See Figure 2)
• Check and replace with the correct fuses if necessary. The fuses
must be 5mm x 20mm quick acting, rated at 250V.
0.5 Cu. Ft. (115VAC)
2.5AF
0.5 Cu. Ft. (230VAC)
1.25AF
1.4 Cu. Ft. (115VAC)
2.5AF
1.4 Cu. Ft. (230VAC)
1.25AF
2.5 Cu. Ft. (115VAC)
2.5AF
2.5 Cu. Ft. (230VAC)
1.25
5.0 Cu. Ft. (230VAC)
3.15AF
• Push the fuse drawer back in. Reconnect unit to power supply.
Fuse
Drawer
Figure 2