System purging – AXEON AT-Series User Manual
Page 19
19
AT – Series User’s Manual
MKTF-205
06/12
SYSTEM PURGING
Carefully inspect your system before initial start-up. Check that all plumbing and
electrical connections are not loose or have not come undone during shipment. A user’s
manual, test results, and filter housing wrench will accompany your AT-Series reverse
osmosis system.
NOTE: LEAVE THE POWER TO THE SYSTEM OFF FOR THIS PROCEDURE.
1. Redirect permeate water to the drain for this procedure.
2. Fully open the concentrate valve #8 (counter clockwise). (Figure 1B, Page 14)
3. Fully close the recycle valve # 7 (clockwise) (if applicable). (Figure 1B, Page 14)
4. Offset the position of the bypass white lever on the solenoid valve #1. (Figure 2,
Page 19)
5. Turn the feed water on and let the system purge until no visible bubbles appear from
concentrate flow meter #11. (Figure 1B, Page 14)
Lever Inline
Lever Offset
Operating Position
Bypass Position
FIGURE 2