beautypg.com

System activation and flushing – AXEON CRO-Series User Manual

Page 8

background image

MKTF - 215

8

System Activation and Flushing

System Activation and Inspection

1. Check all tubing connections to ensure they are firmly seated. CHECK TO SEE THAT THE CARTRIDGE RETAINER CLIP IS

PROPERLY ENGAGED AND LOCKED. Failure to keep the retaining clip in place will result in accidental leaks and flooding.

2. Open the dispensing faucet at the sink. Close the tank shut-off valve. Make sure the feed valve for the incoming water

is open. Open the small feed shut-off valve on the green feed line.

3. Observe all tubing and connections for several minutes to detect any leaks. In approximately 5 minutes, (assuming

normal feed water pressure) the dispensing faucet should begin dripping.

4. Allow the faucet to run for up to 15 minutes, then close the faucet.

5. Check for leaks at all connections.

6. Open the tank shut-off valve.

Initial Flushing Procedure

1. Before the system can be used for drinking water production it must be adequately flushed. Each reservoir tank is

dosed with a small amount of powdered sanitizer before shipment, typically a chlorinating agent, in order to ensure tank

internal cleanliness. Also, the carbon filter cartridge will release a small amount of carbon fines during the first tank full of

flow. This flushing procedure will allow any sanitizer or carbon fines to pass from the system.

2. Initial tank filling will take approximately one hour (based on average feed pressure). When the tank is full, the water

pressure will have risen to the point where the automatic shut-off valve inside the system will stop the feed flow through

the system. Actuation of the automatic shut-off valve can be determined by either checking for a lack of brine flow to the

drain saddle, or by listening very closely near the dispensing faucet for absence of water flow sound though the air gap.

When the tank has filled for the first time, it should be left undisturbed for at least 8 hours to ensure proper sanitization.

3. After 8 hours has elapsed, open the dispensing faucet fully and allow the product water to run out to drain at maximum

flow. The initial discharge will be dark with the bulk of the carbon particle wash out. There may also be the scent of

chlorinated water from the sanitizing agent. When the flow has diminished to a fast drip or small stream, close the

dispensing faucet.

4. Fill and flush the tank at least three times prior to use. If necessary, repeat until the chlorination scent has disappeared.
It is important that the flush be done at maximum flow (e.g. the tank must be full) to assist in rapid wash out. After this
flushing procedure the system is ready for normal use.