Care and Maintenance

STORAGE:

Pocket Gauge: After measurement, fully exhaust cuff then wrap cuff around gauge and

bulb and store in zippered carrying case.

NOTE: This product will maintain the safety and performance characteristics

specified at temperatures ranging from 50°F to 104°F (10°C to 40°C) at a relative

humidity level of 15% to 85%.

Manometer: Your pocket aneroid gauge requires minimal care and maintenance.

The manometer may be cleaned with a soft cloth but should not be dismantled under any cir-

cumstances.

Gauge accuracy can be checked visually; simply be certain

the needle rests within the printed oval when the unit is fully

deflated (Figure 3).

Should the indicator needle of the manometer rest outside of

this calibration mark, then the manometer must be re-cali-

brated to within ±3 mmHg when compared to a reference

device that has been certified to national or international

measurement standards. A manometer whose indicator needle is resting outside of this

mark, is NOT acceptable for use.

In the event that the gauge is ever in need of calibration, simply return for service. Damaged

or broken parts will be replaced as needed at a minimal charge. Refer to the warranty for spe-

cific details of warranty coverage.

The manufacturer recommends a calibration check every 2 years.

Cuff Cleaning and Disinfecting

NOTE: Use one or more of the following methods and allow to air dry:

Wipe with mild detergent and water solution (1:9 solution). Rinse.

Wipe with Enzol per manufacturer’s instructions. Rinse.

Wipe with .5% bleach and water solution. Rinse

Wipe with 70% isopropyl alcohol.

Launder with mild detergent in warm water, normal wash cycle. Remove bladder first.

Cuff is compatible with 5 wash cycles.

Low Level Disinfection

Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray detergent

solution liberally onto cuff and use a sterile brush to agitate the detergent solution over entire

cuff surface for five minutes. Rinse continuously with distilled water for five minutes. To dis-

NOTE: Store gauge with

valve in full exhaust position.