Care and maintenance, Cuff cleaning and disinfecting, Low level disinfection – American Diagnostic Corporation (ADC) Prosphyg 790 User Manual
Page 5
![background image](/manuals/659924/5/background.png)
Care and Maintenance
STORAGE:
Pocket Gauge: After measurement, fully exhaust cuff then wrap cuff around gauge and
bulb and store in zippered carrying case.
NOTE: This product will maintain the safety and performance characteristics
specified at temperatures ranging from 50°F to 104°F (10°C to 40°C) at a relative
humidity level of 15% to 85%.
Manometer: Your pocket aneroid gauge requires minimal care and maintenance.
The manometer may be cleaned with a soft cloth but should not be dismantled under any cir-
cumstances.
Gauge accuracy can be checked visually; simply be certain
the needle rests within the printed oval when the unit is fully
deflated (Figure 3).
Should the indicator needle of the manometer rest outside of
this calibration mark, then the manometer must be re-cali-
brated to within ±3 mmHg when compared to a reference
device that has been certified to national or international
measurement standards. A manometer whose indicator needle is resting outside of this
mark, is NOT acceptable for use.
In the event that the gauge is ever in need of calibration, simply return for service. Damaged
or broken parts will be replaced as needed at a minimal charge. Refer to the warranty for spe-
cific details of warranty coverage.
The manufacturer recommends a calibration check every 2 years.
Cuff Cleaning and Disinfecting
NOTE: Use one or more of the following methods and allow to air dry:
Wipe with mild detergent and water solution (1:9 solution). Rinse.
•
Wipe with Enzol per manufacturer’s instructions. Rinse.
•
Wipe with .5% bleach and water solution. Rinse
•
Wipe with 70% isopropyl alcohol.
•
Launder with mild detergent in warm water, normal wash cycle. Remove bladder first.
•
Cuff is compatible with 5 wash cycles.
Low Level Disinfection
Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray detergent
solution liberally onto cuff and use a sterile brush to agitate the detergent solution over entire
cuff surface for five minutes. Rinse continuously with distilled water for five minutes. To dis-
NOTE: Store gauge with
valve in full exhaust position.
(Figure 3)