beautypg.com

American Diagnostic Corporation (ADC) Prosphyg 790 User Manual

Page 4

background image

pressure. Index and range markings ensure use of

the correct cuff size. The artery mark indicates

proper cuff positioning.

Place the cuff over the bare upper arm with the ar-

tery mark positioned directly over the brachial ar-

tery. The bottom edge of the cuff should be

positioned approximately (1”) one inch (2-3cm)

above the antecubital fold. Wrap the end of the cuff not containing the bladder around the

arm snugly and smoothly and engage adhesive strips (Figure 1).

NOTE: If the unit is equipped with a calibrated

nylon cuff, featuring Index and Range markings, a

correct fit may be verified by checking that the

Index Line falls between the two Range Lines.

(Figure 2).

3. Inflate the cuff

Close the valve by turning thumbscrew clockwise.

Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing

bulb rapidly.

Inflate cuff 20-30 mmHg above the point at which the radial pulse disappears.

NOTE: Cuff pressure range is 0 mmHg to 300 mmHg.

4. Position the Stethoscope

Position the chestpiece in the antecubital space below the cuff, distal to the brachium. Do

not place chestpiece underneath the cuff, as this impedes accurate measurement. Use the

bell side of a combination stethoscope for clearest detection of the low pitched Korotkoff

(pulse) sounds.

5. Deflate the cuff

Open the valve to deflate the cuff gradually at a rate of 2-3 mmHg per second.

6. Measurement

Record the onset of Korotkoff sounds as the systolic pressure, and the disappearance of

these sounds as diastolic pressure. (Some healthcare professionals recommend recording

diastolic 1 and diastolic 2. Diastolic one occurs at phase 4).

NOTE: It is recommended that K4 be used in children aged 3 to 12, and K5 should be used

for pregnant female patients unless sounds are audible with the cuff deflated, in which case

K4 should be used. K5 should be used for all other adult patients.

After measurement is completed, open valve fully to release any remaining air in the cuff.

Remove cuff.

Left Arm

Tube

(Figure 1)

Select Smaller Cuff

CORRECT

Select Larger Cuff