Specifications, 2) output format – A&D Weighing AD-1690 User Manual
Page 24
![background image](/manuals/659069/24/background.png)
22
(2) Output format
- Output of the setting value
When turning the power supply on, the present
setting information is output.
S e t t i n g
I n f o .
CR LF
0 0 . 3 k P a / 0 3 . 0 s
CR
LF
When changing the setting value, the new setting information is output.
S e t t i n g
C h a n g
e d
CR LF
0 0 . 3 k P a / 0 3 . 0 s
CR
LF
- Output of the measurement result
When finishing the measurement, the data is output one time.
When PASS: the judgment result“PASS”is output and the measurement pressure value.
P A S S
- 1 9 . 9 k
P a
CR
LF
When FAIL: the judgment result“FAIL”is output and the measurement pressure value.
F A I L
- 1 8 . 5 k
P a
CR LF
Or
F A I L
L o w
CR
LF
When not evacuating the air to –20kPa caused by many leaks.
12. SPECIFICATIONS
Pressure of
evacuating the air
-
20kPa ± 20% (Fixed)
Capacity pressure
changing of Inside
instrument
+0.2kPa / 10 seconds
Setting value of quantity of pressure change:
+0.1~+20kPa (Minimum change value: 0.1kPa, variable)
Leak judgment
condition
Monitor time:
1.0 second~ (Minimum change value: 0.5 second, variable)
Pump driving time ~6 seconds
Dimensions (W)
231mm×(D) 126mm×(H) 78mm
Weight of the
main unit
Approx. 570g
Attachment
(Large) For micro pipette with more than 5000μℓ to less than 10000μℓ
(Medium) For micro pipette with 1000μℓ
(Small) For micro pipette with more than 20μℓ to less than 200μℓ
(Smallest) For micro pipettes of less that 10μℓ
* When your pipette and the attachment tip size do not match up, replace
the attachment tip with a suitable sized tip.