Aout – auto print enable when accumulate weight met.
•
Beep – beep tone for key press
•
C-fact - conversion factor for custom units. Will multiply weights, in primary units, to
get custom units. Divide primary units by C-factor to get display in custom units.
Cal Menu
•
SL – security level – see manual for complete setup information.
•
Dl - (deal load) flashes dl then goes to raw counts for condition of scale input
•
Calat 1 – reference weight value for first calibration point (used in both single point or
multiple point calibration.
•
Calat 2 – 4 – reference weight values for second to forth calibration points in a multiple
point calibration.
•
Load 1-4 - prompt to place test load on scale for each cal point. Will show raw counts
for weights.
Calibrating the scale
1. Press and hold Program/Enter key until “Setup” appears
2. Press Program/Enter to enter the Setup menu.
1. Set the parameters in setup menu. These settings will determine how the
unit will handle the serial ports.
3. Press the Over/under/next key to go to “Config”
4. Press Program/Enter to enter the config menu.
1. Set the parameters in the config menu. These settings will determine
how the unit will handle the weight readings and how it responds to
changes in weight and tolerances used.
5. Press the Over/under/next key to go to “APP”
6. Press Program/Enter to enter the APP menu.
1. Set the parameters in the App menu. These settings will determine how
the unit will handle the printer outputs. Also sets up the C-factor for
custom units.
7. Press the Over/under/next key to go to “Cal”