O.r – zero tolerance range (set as % of capacity)
o
100
o
2
•
D – division size
o
50, 20, 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001,
0.0005, 0.0002, 0.0001
•
Filter – (intended to minimize effects of motion, vibration, and wind currents)
•
Flf – Flush filter. Allows switch to fast filter rate if weight changes quickly
•
Tare – enable tare function
o
Off – tare inactive
o
On – tare is active
o
On-Clr – automatically clears when gross wt returns to zero.
•
Hl – overload limit (set as % of capacity)
o
102.5
o
103
o
105
o
110
o
300
•
Ul – underload limit (set as % of capacity below zero)
o
5.5
o
10
o
25
o
50
o
100
•
D rate – display update rate (set in tenths of seconds)
o
From 0.1 to 1.0
•
P1 – serial port 1 output format
o
GTN – if in gross mode will print gross weight. If in net mode, will print
gross/tare/net weights
o
nEt – will print net only
•
P2 – serial port 2 output format
o
GTN
o
nEt
APP menu
•
X.xxupd – display of micro volts per division. Should be greater than 1.0. Read only.
If less than 1.0, adjust division size.
•
Acc – Enable/disable accumulator
•
A thld – Autoprint and accumulator threshold. Number of divisions to meet or exceed to
trigger auto print or accumulator.