beautypg.com

8 maintenance, Maintenance, 8maintenance – FRITSCH PULVERISETTE 23 User Manual

Page 27

background image

8

Maintenance

DANGER!
Mains voltage
– Before beginning with maintenance work, unplug the

mains plug and protect the device against being unin-

tentionally switched back on again!

– Indicate maintenance work with warning signs.
– Maintenance work may only be performed by special-

ised personnel.

– Put safety equipment back into operation after mainte-

nance or repair work.

We recommend keeping a safety logbook Ä Chapter 13

‘Safety logbook’ on page 35, where all work (maintenance,

repairs......) performed on the device is entered.

The most important element of maintenance is regular

cleaning:

When cleaning the complete device, adhere to the

guidelines of the Accident Prevention Regulation (BGV

A3) ‐ especially if the device has been set up in a dusty

environment or if the processed source material pro‐

duces dust.

n

In stainless grinding bowls a burr can form over time on the inner

edge between the bowl and the lid due to the ductility of the mate-

rial. This burr causes the bowl and the lid to rest on each other,

metal against metal, and the O-ring seal ceases to function. If you

notice burr formation, it should be removed immediately with a

scraper or fine emery paper before the burr becomes so large that

removing it is difficult and requires a lot of effort.

Functional part

Task

Test

Maintenance interval

Felt ring as piston seal

Sealing the piston bearing

Put a few drops of oil on

the piston

Every 50 operating hours.

Drive motor

Permanent lubrication

Bearing clearance

Every 400 h or annually

Connection rod bearing

Permanent lubrication

Bearing clearance

Every 400 h or annually

Piston bearing

Maintenance free slide

bearing

Bearing clearance

Every 400 h or annually

Spindle for the grinding

bowl clamping device

Clamping the grinding

bowls

Mobility, oil if necessary

Every 100 h

Maintenance