The Compex Wireless stimulator needs special precautions regarding EMC and needs to be installed and put into service according to the EMC
information provided.
Portable and mobile RF communications equipment can affect the Compex Wireless stimulator.
The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result in increased Emissions or decreased
Immunity of the Compex Wireless stimulator.
The Wireless stimulator should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary, the Compex
Wireless stimulator should be observed to verify normal operation in the configuration in which it will be used.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Compex Wireless stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the
Compex Wireless stimulator should assure that it is used in such an environment.
EMISSION TESTS
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Radiated Emissions CISPR 11
Group 1
The Compex Wireless uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Conducted Emissions CISPR 11
Class B
The Compex Wireless is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations IEC 61000-
3-3
Complies
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Wireless stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless
stimulator should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for
input/output lines
±2kV (power lines)
Not Applicable (I/O lines)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1kV Line to Line
(or Neutral)
±2kV line(s) to earth
±1kV Line to Line
Not Applicable
(Line to Earth)
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11 a
<5% UT (>95% dip in UT) for 0,5
cycle
40% UT (60% dip in UT) for 5
cycles
70% UT (30% dip in UT) for 25
cycles
<5% UT (>95% dip in UT) for 5 sec
<5% UT (>95% dip in UT) for 0,5
cycle
40% UT (60% dip in UT) for 5
cycles
70% UT (30% dip in UT) for 25
cycles
<5% UT (>95% dip in UT) for 5 sec
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the Compex Wireless
stimulator requires continued
operation during power mains
interruptions, it is recommended
that the Compex Wireless
stimulator be powered from an
uninterrupted power supply or
a battery.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
a
U
T
is the a.c mains voltage prior to application of the test level.