8 technical specifications – ChemoMetec SCC-100 User Manual
Page 47

8 Technical specifications
35
8 Technical specifications
8.1
Technical specifications for NucleoCounter SCC-100
Specificity
Somatic cells stained with the DNA specific fluorescent dye, PI.
Sample types
Milk samples containing dispersed somatic cells.
Sample consumption
The sample is mixed with a lysis buffer prior to analysis. It is
recommended that at least 500 µl of sample are used for this
dilution in order to reduce error caused by sample handling.
Approximately 50 µl of the lyzate mixture is loaded into the
cassette
Analysis volume
Approximately 2 µl of the lyzate mixture is being analyzed in
the NucleoCounter SCC-100, corresponding to approximately 1
µl of milk.
Somatic Cell Count in a volume of milk sample
Measurement range
Total range:
Recommended range:
10,000 to 2,000,000 cells/ml
100,000 to 750,000cells/ml
Presentation of Result
The NucleoCounter SCC-100 presents the estimated SCC in the
milk sample when milk and Reagent C are mixed in equal
amounts.
Operation
Menu-controlled by means of keyboard and LCD display.
Analysis time
When pressing “Run” on the NucleoCounter SCC-100 the result
will be displayed within 30 seconds.
Physical data
Weight
Height
Width
Depth
3 kg
26 cm
38 cm
22 cm
Power
See Section 0 Power Supply
Input
Fuse
12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)
Power consumption
NucleoCounter
Peak
Ready mode
Standby
25 W
2.5 W
2 mW
Operation conditions
Maximum relative humidity 80 percent for temperatures up to
31°C decreasing linearly to 65 percent relative humidity at
maximum 35°C; minimum temperature 15°C.
USB
USB, version 1.1.
Note: Does not support USB Hubs.
Printer
Thermal printer with RS-232 interface and cable optionally
supplied by ChemoMetec A/S.