beautypg.com

User's manual for opto-rs cable connections – Checkline J-CC User Manual

Page 2

background image

User's manual for OPTO-RS cable connections

General

The OPTO-RS cable enables a direct connection with most of the Käfer measuring instruments to a personal
computer, a dedicated printer or to a Käfer display unit.

It is not only a cable, but an interface which converts the data output of the instrument to a compatible RS232
signal. The periphery instrument connection must be able to supply power to the OPTO-RS plug.


Definition

RS232 communication parameters

4800 baudrate, even parity, 7 data bits, 2 stop bits


Data format

Data

[ Sign ¦ E1-En ¦ "." ¦ F1-Fn ¦ CR ] Sign : « + », « - », or « space »

E1-En: integer
F1-Fn: fractional
n: depends on used unit and resolution

Errors

[ "ERR" ¦ Number ¦ CR ]

0: sensor error (e.g. speed, scale distance)

1: incorrect command

2: parity error (duplex instruments only)

3: exceeded measurement range

Identification

[ "SY" ¦ Instr. ¦ "." ¦ OPT1 ¦ {"." ¦ OPT2} ¦ CR ]

SY: Sylvac Instr: 203, 235, 233, etc OPT1: version option OPT2: additional version options (according to
instrument used) Note: The id. transmission is done only when switching ON the instrument


Connection description