Scan

Normally, the PosiTector UTG takes a single spot measurement
at a rate of 6 readings per second while in contact with a surface.
When the probe is lifted, the last reading will remain on the display.

It is sometimes necessary to examine a larger region to locate the
thinnest point. When Scan is selected, the PosiTector UTG will
measure at a rate of 20 readings per second and display min/max
values.

Reset

Reset

(soft reset) restores factory settings and returns the

instrument to a known condition. The following occurs:

- All batches, stored measurements, images, and batch names

are erased.

- All calibration adjustments are cleared and returned to the

Gage’s factory calibration settings.

- Menu settings are returned to the following:

Memory = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF

A Scan = OFF
B Scan = OFF
Cal Lock = OFF
Bluetooth = OFF

Perform a more thorough Hard Reset by powering down the
instrument, waiting several seconds, then simultaneously holding
both the center and (+) buttons until the Reset symbol
appears. This returns the instrument to a known, “out-of-the-box”
condition. It performs the same function as a menu Reset with the
addition of:

- Date and Time are not affected by either Reset.

Units = millimeter
Flip Display = Normal
Auto Sync = OFF
White on Black = OFF
Language = English

Battery Type = Alkaline
Backlight = Normal
Bluetooth Streaming = OFF
USB Drive= ON

- Bluetooth Pairing info is cleared.

- Menu settings are returned to the following:

Setup Menu

Setup Menu

NOTE:

Display = None
SE Mode = OFF

8