3 hose connections for the triplescrubeco, 3 hose connections for the triplescrub – BUCHI Scrubber K-415 User Manual
Page 34
5 Putting into operation
34
K-415 Operation Manual, Version B
5 .7 .3
Hose connections for the TripleScrub
ECO
The TripleScrub
ECO
consists of the neutralization, the adsorption vessel, the condenser unit and the
cooling water control.
To establish the hose connections, proceed as follows:
• Put hose connectors GL 14 with seals on all PTFE hoses.
• Connect the emission source to the condenser inlet
g
.
• Connect the condenser outlet
h
to the neutralization vessel
a
.
• Connect the adsorption vessel
b
to the bypass valve
e
.
• Connect your cooling water source to the cooling water tube of the water control
c
using the
provided adapter with sieve and tighten it (see chapter 5.6).
• Cut off a piece of the provided silicone hose and connect the cooling water outlet
d
of the water
control
c
to the cooling water inlet
i
of the condenser. Secure both connections with hose clamps.
• Cut off a piece of the provided silicone hose and connect the cooling water outlet
j
of the
condenser
k
to the waste water system.
3
6
4
5
9
8
11
1
2
10
7
a Neutralization stage
b Adsorption vessel
c Water control
d Water control (outlet)
e Bypass valve
f Silencer
g Inlet from emission source (digester)
h Outlet to neutralization vessel (top)
i Cooling water inlet (bottom)
j Cooling water outlet (top)
k Condenser
Fig. 5.13: Hose connections for the TripleScrub
ECO