2 technical data overview, 3 materials and approvals, Technical data overview – BUCHI Scrubber K-415 User Manual
Page 18: Materials and approvals, 3 .2 technical data overview, 3 .3 materials and approvals
3 Technical data
18
K-415 Operation Manual, Version B
3 .2
Technical data overview
Table 3-7: Technical data
Scrubber Unit K-415
Dimensions (W x H x D)
(completely equipped instrument)
DuoScrub
TripleScrub
TripleScrub
ECO
QuadScrub
ECO
248 x 412 x 526 mm
347 x 417 x 526 mm
347 x 417 x 543 mm
347 x 450 x 543 mm
Weight
DuoScrub
TripleScrub
TripleScrub
ECO
QuadScrub
ECO
10.1 kg
11.2 kg
12.3 kg
12.9 kg
Power system voltage / frequency /consumption / fuse 230 V ± 10 % / 50/60 Hz
120 V ± 10 % / 50/60 Hz
100 V ± 10 % / 50/60 Hz
140 W
T1A L 250 V
140 W
T2A L 250 V
140 W
T2A L 250 V
Installation site
For indoor use only
Temperature
+ 5 °C to + 40 °C
Altitude
up to 2000 m above sea level
Humidity
maximum relative humidity 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Pump suction capacity (max.)
32 L/min (air)
System suction capacity
adjustable approximately between 50 and 400 mbar
below atomspheric pressure
Overvoltage category
II
Pollution degree
2
Noise level
< 70 dB
3 .3
Materials and Approvals
Table 3-8: Materials and Approvals
Component
Material designation
Housing
PUR
Minor hardware
PP, PA, POM
Lip gasket acid- / base-resistant
FPM, EPDM
Gasket
PTFE, CR
Cover of neutralization vessel
PMMA
Glass parts
Borosilicate 3.3, P+G plastic coated glass
Pump
PPS, CR, FPM
Hoses
CR, FEP
Cooling water control
brass, FKM (sealing)
Water hose (black)
Polychloroprene Rubber
Water hode (white)
brass, PA12
Approvals
CE, CSA