2 hose connections for the triplescrub – BUCHI Scrubber K-415 User Manual
Page 33
5 Putting into operation
33
K-415 Operation Manual, Version B
5 .7 .2
Hose connections for the TripleScrub
The TripleScrub consists of the neutralization vessel, the adsorption vessel, and the condenser unit.
To establish the hose connections, proceed as follows:
• Put hose connectors GL 14 with seals on all PTFE hoses.
• Connect the emission source to the condenser inlet
e
.
• Connect the condenser outlet
f
to the neutralization vessel
a
.
• Connect the adsorption vessel
b
to the bypass valve
c
.
• Cut off a piece of the provided silicone hose and connect the cooling water source to the cooling
water inlet of the condenser
g
. Secure it with a hose clamp.
• Cut off a piece of the provided silicone hose and connect the cooling water outlet of the condenser
h
to the waste water system.
1
2
3
4
5 6 7
8
9
a Neutralization stage
b Adsorption vessel
c Bypass valve
d Silencer
e Inlet for fumes from emission source
f Outlet to neutralization vessel (top)
g Cooling water inlet (bottom)
h Cooling water outlet (top)
i Condenser
Fig. 5.12: Hose connections for the TripleScrub