6 operation 27 – BUCHI Melting Point M-565 User Manual
Page 27

6 Operation
27
MeltingPoint Monitor 1.1 Operation Manual, Version B
6.6.2
System Information report
A system information report contains all settings and service information of a Melting Point M-565. If
requested, use this report to provide detailed information to the customer service.
Link the Melting Point M-565 to the computer and select Tools > Get SysInfo, to generate a PDF.
Settings
Service
Language
english
Serial number
12.12.12.1
Stop after detection
enabled
Firmware version
1.0.0
Param adjust. during Measurement
disabled
CPLD version
0.0
Result printout automaically
disabled
Operating hours
320
Result graph indication
enabled
No. of melting point measurements
55
Calibration interval
12 Months
No. of boiling point measurements
15
Sample ID Request
enabled
No. of calibrations
0
Barometric pressure request
never
Last calibration
1/28/2009
Boiling sample conditioning
30 s
Last verification
10/30/2008
Melting result
thermodynamic
Factory adjustment at 20°C
0.00 °C
Thermodynamic factor
1.5
Calibration value at 52 °C
0.00 °C
Company name
MARCEL
Calibration value at 148 °C
0.00 °C
Date (DD.MM.YYYY)
2/2/2009
Calibration value at 237 °C
0.00 °C
Time (24h)
12:06 PM
Calibration value at 335 °C
0.00 °C
Date format
DD MM YYYY
Factory adjustment at 400 °C
0.00 °C
Time format
european (24h)
Sample illumination brightness
50 %
Temperature unit
°C
Camera position horizontally
20
Pressure unit
mmHg
Camera position vertically
20
Buzzer volume
0
Color gain red for melting point
87
External keyboard
EN_US
Color gain green for melting point
52
Display brightness
100 %
Color gain blue for melting point
38
Color gain red for boiling point
20
Color gain green for boiling point
18
Color gain blue for boiling point
77
Date of board test
1/1/0001
Date of end test
1/1/0001
Board temperature
26.0 °C
Heating present
OK
Cooling fan present
OK
Housing fan present
OK
Temperature sensor present
OK
Temperature sensor function
OK
24V after fuse present
OK
24V input present
OK
5V present
OK
M-565
CompanyName
SysInfo
MeltingPoint Monitor V1.0
Date, signature
Sysinfo.pdf
1 / 1
6.6.3
Test Result report
A ‘Test Result’ report can be generated after performing a hardware test at the Melting Point M-565.
It contains information about the hardware functionality only. If requested, use this report to provide
detailed information to the customer service.
Link the Melting Point M-565 to the computer and select Tools > Get Test Report, to generate a PDF.
Test Result
Keypad
OK
Camera value without sample L
13
Rotary knob
OK
Camera value without sample C
11
24 V input present
OK
Camera value without sample R
11
24 V after fuse present
OK
Camera value with sample L
49
5 V present
OK
Camera value with sample C
57
Heating present
OK
Camera value with sample R
80
Cooling fan present
OK
Housing fan present
OK
Temperature sensor present
OK
Temperature sensor function
OK
External keyboard
Not tested
Board temp. sensor
OK
Heating function
OK
Cooling fan
OK
Housing fan
OK
Sample illumination
OK
Camera position
OK
Camera brightness
Not OK
Internal clock
OK
Display
Not OK
M-565
CompanyName
Test
MeltingPoint Monitor V1.0
Date, signature
DeviceTest1212009pdf.pdf
1 / 1