6 system generated reports, 6 operation 26 – BUCHI Melting Point M-565 User Manual
Page 26

6 Operation
26
MeltingPoint Monitor 1.1 Operation Manual, Version B
6.6
System generated reports
6.6.1
Calibration report
After a successful system calibration, the software will automatically generate a report in PDF format
if the PC is linked. In case the link has not been established, the report is stored in the temporary
memory and can be received via Tools > Get Latest Calibration.
NOTE
During calibration, viewing existing results or changing methods stored on the computer is
•
possible.
The calibration result stays available on the Melting Point M-565 until it is switched-off or a new
•
determination is started!
Temp. gradient
0.5 °C/min
Standard deviation limit
0.2 °C
No. of required samples
6
4-Nitrotoluene
Lot: 1234567890
Automatic:
Left
Center
Right
1:
50.1 °C
52.5 °C
52.3 °C
2:
54.9 °C
52.5 °C
3:
52.6 °C
53.7 °C
4:
52.8 °C
52.4 °C
56.0 °C
Average
52.5 °C
Standard deviation
0.17 °C
Reference temperature
52.0 °C
Diphenylacetic Acid
Lot: Subst. 2
Automatic:
Left
Center
Right
1:
148.1 °C
148.2 °C
148.3 °C
2:
148.9 °C
148.5 °C
3:
148.6 °C
148.7 °C
4:
148.8 °C
148.4 °C
148.0 °C
Average
148.3 °C
Standard deviation
0.19 °C
Reference temperature
148.0 °C
Caffeine
Lot: Coffein
Automatic:
Left
Center
Right
1:
237.1 °C
237.2 °C
237.3 °C
2:
237.9 °C
237.5 °C
3:
237.6 °C
237.7 °C
4:
237.8 °C
237.4 °C
237.0 °C
Average
237.3 °C
Standard deviation
0.19 °C
Reference temperature
237.0 °C
Potassium Nitrate
Lot: Kalibriersubstanz 4
Automatic:
Left
Center
Right
1:
335.1 °C
335.2 °C
335.3 °C
2:
335.9 °C
335.5 °C
3:
335.6 °C
335.7 °C
4:
335.5 °C
335.4 °C
335.0 °C
Average
335.5 °C
Standard deviation
0.14 °C
Reference temperature
335.0 °C
Last calibration
1/16/2009
Serial number
12.12.12.1
Last verification
10/30/2008
Firmware version
0.39.1
Date
1/16/2009
Time
9:54 AM
M-565
CompanyName
Calibration
MeltingPoint Monitor V1.0
Date, signature
Calibration1162009.pdf
1 / 1